#21 Re: stacja GM330 (PC410 i REX C-100) odmawia posłuszeństwa


by matic 25 April 2018, 22:39
Your "STOP" button wiring is correct, but your "START" button wiring is wrong. The "START" button should be NC (Normally Closed) type (short circuit in the initial state and open circuit, when it is pressed). You should just connect one wire to another contact of the START button. Post the picture of the back side of your START button and I could tell you how to connect it.

Correct values:
Resistance between pins 13 and 15 of PC410 controller in the following cases:
a) STOP button released -> Infinite
b) STOP button pressed -> Cose to 0Ω - short circuit (2-4Ω should be also ok)

Resistance between pins 14 and 15 of PC410 controller in the following cases:
c) START button released -> Cose to 0Ω - short circuit
d) START button pressed -> Infinite

radio790 wrote:Czy mogło to coś uszkodzić??
Probably not, if the machine was disconnected from the mains socket.

Re: Re: stacja GM330 (PC410 i REX C-100) odmawia posłuszeństwa


by Google Adsense [BOT] 25 April 2018, 22:39

#22 Re: stacja GM330 (PC410 i REX C-100) odmawia posłuszeństwa


by radio790 26 April 2018, 02:24
Powiedzmy że rozumiem ale jak mogą być złe podłączenia skoro 5 lat stacja działała 100% OK.
Nikt tam nie grzebał. Problemy zaczęły się po wgraniu profili programem BGA Maintain Station lub klikanie dziecka.

Image

Stacja działa podłączona do komputera ale jak widać na zdjęciach z wcześniejszego postu pomiary temperatury i czasy szaleją.

Wyciągnięty PC410 z maszyny też się źle zachowuje bo piszczy itd.

może na nowo ustawić PID? tylko jak?

#23 Re: stacja GM330 (PC410 i REX C-100) odmawia posłuszeństwa


by matic 26 April 2018, 07:40
In my rework station (Achi IR PRO SC V3) the "START" button is "NC" type. The only difference is that, your PC410 controller is probably from another manufacturer. Mine is original Altec, yours is Shanm.

radio790 wrote:Po naciśnięciu START bez reakcji. Gdy trzymam START stacja rusza bez podgrzewacza dolnego i po puszczeniu zatrzymuje się i utrzymuję temperaturę górnego.
I think you have nothing to loose, if you try to connect the "START" button as I said.
Maybe this will solve that issue.

Just move both blue wires from central contact to the top one.
Image

I'm pretty sure there is no way to reset the "PID" settings to the factory default values.
There are "Auto tune" option, but I don't know if it works well or not, because I have never used it.

#24 Re: stacja GM330 (PC410 i REX C-100) odmawia posłuszeństwa


by radio790 26 April 2018, 12:08
Matic jest postęp!!! zmiana styku pomogła :) górny podgrzewacz działa ok (test bez komputera) temperatury OK czasy OK.

Niestety dolny podgrzewacz nadal nie startuje (dolny nie działa).

Może teraz to automatyczne strojenie?? chyba ze masz inny pomysł jak zachęcić dolny podgrzewacz do pracy ?

Kolego Matic właśnie wykonałem test na komputerze (zmieniłem dodatkowo adapter USB-RS232 z Prolific PL2303 na FTDI Digitus i na nowo wgrane profile)
Wszystko wydaje się być ok co widać na fotkach. Dolny podgrzewacz działa z tym ze przez pierwsze 3 etapy temperatura większa o 18°C (ustawione 80°C dało 98°) kolejne etapy OK (ustawione 80°C i trzymało 80°C). Jak poprawić dolny podgrzewacz? Termopara ustawiona przez producenta do grzałek chyba ze to nie jest problem.

Image
Image



Test bez komputera: nadal dolny podgrzewacz nie działa (nie startuje)
liczę Matic na Twoją wiedzę jak to poprawić ?

#25 Re: stacja GM330 (PC410 i REX C-100) odmawia posłuszeństwa


by matic 27 April 2018, 21:39
I think that the "PID" settings has nothing to do with the lower heater.

Remove the PC410 controller from the machine, solder the ON-OFF switch between highlighted pins on the following picture (do not solder to the gold contacts!!!).

1) Put the switch to ON position.
2) Connect the 230V to pins 1 and 2 of PC410.
3) To simulate the "START" button put the switch to OFF position and immediately back to ON.
4) If the PC410 starts, measure the voltage to ground on highlighted pin before and after simulating the "START" button.

Image

#26 Re: stacja GM330 (PC410 i REX C-100) odmawia posłuszeństwa


by radio790 4 May 2018, 15:46
Matic nie mogę wykonać tego zabiegu ponieważ jak wcześniej po podłączeniu bezpośrednio pod 230v brzęczyk wyje i na ekranie to co wcześniej.

Tak po za tym jest OK. Nadal kombinuje coś aby w trybie manualnym działał dolny podgrzewacz ale brakuje mi pomysłu.

Dziwi mnie fakt ze przez Komputer wszystko działa.

#27 Re: stacja GM330 (PC410 i REX C-100) odmawia posłuszeństwa


by matic 5 May 2018, 09:36
radio790 wrote:Matic can not perform this procedure because as before, after connecting directly to 230v, the buzzer buzzes and on the screen what it was before.
Does the controller still starts the profile itself if you bridge the pins, which I highlighted in post #25 and then connect it to 230V?

#28 Re: stacja GM330 (PC410 i REX C-100) odmawia posłuszeństwa


by radio790 6 May 2018, 10:52
!!!!SUKCES!!!!

zmieniłem ustawienia PC410 :
ALO1-----OFF
ALO2------Pout

W trybie manualnym (bez komputera) wszystko działa :)
pod komputerem wszystko działa :)Temat można uznać za zamknięty.

chyba że walczymy jeszcze z zawyżoną temperaturą podgrzewacza dolnego o 15°C? (idzie się przyzwyczaić biorąc na to poprawkę)

dzięki za pomoc

ps matic. co się stało ze stacją (co było nie tak) ze zmiana pin START który zmieniłem na twoje polecenie ??

#29 Re: stacja GM330 (PC410 i REX C-100) odmawia posłuszeństwa


by matic 6 May 2018, 13:23
radio790 wrote:zmieniłem ustawienia PC410 :
ALO1-----OFF
With this setting you disabled the sound alarm when the profile ends.
If you want to turn it back ON set "ALO1 --> P.End".

radio790 wrote:ALO2------Pout
This is correct setting.

radio790 wrote:chyba że walczymy jeszcze z zawyżoną temperaturą podgrzewacza dolnego o 15°C?
Can you describe this issue in more detail please?

radio790 wrote:ps matic. co się stało ze stacją (co było nie tak) ze zmiana pin START który zmieniłem na twoje polecenie ??
As I know, the "Start" button on these controllers should be NC (Normally Closed) type for normal operation. It usually works with NO (Normally Open) type as well, but sometimes it happens the strange things in this case.
I don't know, why the "Start" button was connected as NO (Normally Open) type in your machine. Perhaps this was a mistake in assembling the machine in factory.

#30 Re: stacja GM330 (PC410 i REX C-100) odmawia posłuszeństwa


by radio790 6 May 2018, 19:56
A wiec jeżeli chodzi o zawyżona temperaturę to zarówno na komputerze jak i manualnie przez 3-4 etapy profilu temperatura jest wyższy niż ustawiona o 15°C tzn. jak ustawie 180 to przez te 3 etapy grzeje dolna na 195 po tych 3-4 etapach profilu spada do ustawionej 180. I jaką bym temperaturę ustawił zawsze te magiczne 15-16°C przez pierwsze etapy. jak ustawie 100 to grzeje 115. Nie wiem czy to normalne zachowanie ale wcześniej tak nie miałem.

Re: stacja GM330 (PC410 i REX C-100) odmawia posłuszeństwa


by Google Adsense [BOT] 6 May 2018, 19:56
This topic is marked as ARCHIVAL. Please only reply if your message contains the solution (Terms and Conditions p. 12.1).

Who is online

Users browsing this forum: No registered users and 7 guests

_______________________________
All rights reserved. Unauthorised copying of this website's content or any of its part is strictly forbidden.
Any trademarks, brand names, products or services published on this website belong to their legal owners, are copyrighted and used for information purposes only.